Lake Shore Model 475 Gaussmeter User’s Manual
Remote Operation
6-31
BRIGT?
Display Brightness Query
Input:
BRIGT? [term]
Returned:
<bright>[term]
Format:
n
(Refer to command for description)
CMODE
Field Control Mode Command
Input:
CMODE <mode>[term]
Format:
n
<mode>
Specifies the control mode. Valid entries: 0 = Off, 1 = Closed Loop PI.
Example:
CMODE 1[term]
– Field Control uses Closed Loop control.
CMODE?
Field Control Mode Query
Input:
CMODE? [term]
Returned:
<mode>[term]
Format:
n
(Refer to command for description)
CPARAM
Field Control Parameter Command
Input:
CPARAM <P value>, <I value>, <ramp rate>, <control slope limit>[term]
Format:
+nnn.nnnE±nn,+nnn.nnnE±nn,+nnn.nnnE
±
nn,+nnn.nnnE
±
nn
<P value>
The value for control loop Proportional (gain): 0.01 to 1000.
<I value>
The value for control loop Integral (reset): 0.0001 to 1000
<ramp rate>
Field Setpoint ramp in Units/minute, where Units are the present Field
measurement units. Setting to 0 will turn ramping off.
<control slope limit> Analog Output limit in Volts/minute: 0.01 to 1000.
Example:
CPARAM 10,50,3000,75[term]
– Field Control P is 10, I is 50 and when the
CSETP
command is
issued, the field setpoint will ramp at a rate of 3000 Gauss/minute. The Analog Output will be limited
to a 75 Volts/minute change.
CPARAM?
Field Control Parameter Query
Input:
CPARAM? [term]
Returned:
<P value>, <I value>, <ramp rate>, <control slope limit>
Format:
±
nnnnnn,
±
nnnnnn,
±
nnn.nnnE
±
nn,
±
nnn.nnnE
±
nn
(Refer to command for description)
CSETP
Field Control Setpoint Command
Input:
CSETP <value>[term]
Format:
±
nnn.nnnE
±
nn
<value>
Specifies the final setpoint of the Field Control Ramp:
±
350 kG
Example:
CSETP 3000[term]
– Ramp the field setpoint to 3000 (resent units) based on parameters configured in
the command
CPARAM
.
CSETP?
Field Control Present Setpoint Query
Input:
CSETP? [term]
Returned:
<setpoint> [term]
Format:
±
nnn.nnnE
±
nn
(Refer to command for description)
Remarks:
Returns the present field setpoint. If the setpoint is ramping this will be the value where the ramp is,
not the final destination setting.
Содержание 475
Страница 16: ...Lake Shore Model 475 Gaussmeter User s Manual 1 6 Introduction 1 4 SAFETY SYMBOLS ...
Страница 25: ...Lake Shore Model 475 Gaussmeter User s Manual Background 2 9 Figure 2 8 Examples of PI Control ...
Страница 36: ...Lake Shore Model 475 Gaussmeter User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 475 Gaussmeter User s Manual 4 32 Operation This Page Intentionally Left Blank ...
Страница 82: ...Lake Shore Model 475 Gaussmeter User s Manual 6 6 Remote Operation Figure 6 2 Model 475 Status System ...