Lake Shore Model 475 Gaussmeter User’s Manual
5-2
Advanced Operation
Some probes also feature temperature compensation. Turning Temperature Compensation OFF causes the Model 475 to
ignore this data.
To configure the Field Compensation setting, press the Probe key and wait until the Probe Serial number display
disappears and the following screen appears as a prompt for field compensation.
Sel ec t Wi t h ° ®
Fi e l d Co mp Of f
Use the
s
or
t
key to select Off or On. Press
Enter
to accept the new selection and continue to the next setting screen.
Press
Escape
to cancel the selection and return to the normal display.
The next probe setting screen appears as a prompt for changing the temperature compensation setting.
Sel ec t Wi t h ° ®
Te mp Co mp Of f
Use the
s
or
t
key to select Off or On. Press
Enter
to accept the new selection and return to the normal display. Press
Escape
to cancel the selection and return to the normal display. If the probe does not include temperature compensation,
this setting is ignored.
5.1.4
Extension Cable
The complex nature of Hall-effect measurements makes it necessary to match extension cables to the probe when longer
cables are needed. Keeping probes and their extensions from getting mixed up can become a problem when more than
one probe is in use. The Model 475 alleviates some of the difficulty by allowing users to match probes to extensions in
the field. Stored information can be viewed on the front panel and read over computer interface to ensure proper mating.
NOTE:
Calibrated probe and extension cable pairs may be purchased from Lake Shore and are recommended for the
best accuracy. Extension cables with no calibration may also be used. The Model 475 can then be used to match a probe
to the un-calibrated extension cable.
To match a probe with an extension cable, verify that the probe is connected to the instrument. To begin the matching
process,
press and hold
the
Probe
key for approximately 4 seconds. The following screen appears as a prompt for
selecting the programming method.
Sel ec t Wi t h ° ®
MPEC Pr o g r a m
Use the s or t key to select from MPEC program (extension cables) or MCBL Program (User programmable cables,
refer to Paragraph 5.2.1). Press
Enter
to accept the new selection and continue the matching process. Press
Escape
to
cancel the selection and return to the normal display.
The next probe matching screen appears as a prompt to begin the copying process.
At t ac h MPEC c abl e
En t e r t o Co n t i n u e
Содержание 475
Страница 16: ...Lake Shore Model 475 Gaussmeter User s Manual 1 6 Introduction 1 4 SAFETY SYMBOLS ...
Страница 25: ...Lake Shore Model 475 Gaussmeter User s Manual Background 2 9 Figure 2 8 Examples of PI Control ...
Страница 36: ...Lake Shore Model 475 Gaussmeter User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 475 Gaussmeter User s Manual 4 32 Operation This Page Intentionally Left Blank ...
Страница 82: ...Lake Shore Model 475 Gaussmeter User s Manual 6 6 Remote Operation Figure 6 2 Model 475 Status System ...