Lake Shore Model 475 Gaussmeter User’s Manual
4-6
DC Operation
Operation
4.6
DC MEASUREMENT MODE
To measure static or slowly changing fields, press the
DC
key on the front panel. In DC measurement mode, the keypad
and functionality of the Model 475 is optimized to provide the best interaction for DC measurements. The keypad
features are described in paragraphs 4.6.1 through 4.6.7.
4.6.1
DC Operation Select Range and Autorange
The Model 475 reads each Lake Shore probe type: High Stability, High Sensitivity, and Ultra-High Sensitivity. The
tables below list full scale ranges for each probe sensitivity, along with the
maximum number of digits of display
resolution
. Measurement resolution (noise floor) varies depending on probe and application.
High Stability Probe (HST)
Gauss
Tesla
Oersted
Amp/meter
Range and Resolution
Range and Resolution
Range and Resolution
Range and Resolution
±350.000 kG
±35.0000 T
±350.000 kOe
±28.0000 MA/m
±35.0000 kG
±3.50000 T
±35.0000 kOe
±2.80000 MA/m
±3.50000 kG
±350.000 mT
±3.50000 kOe
±280.000 kA/m
±350.000 G
±35.0000 mT
±350.000 Oe
±28.0000 kA/m
±35.0000 G
±3.50000 mT
±35.0000 Oe
±2.80000 kA/m
High Sensitivity Probe (HSE)
Gauss
Tesla
Oersted
Amp/meter
Range and Resolution
Range and Resolution
Range and Resolution
Range and Resolution
±35.0000 kG
±3.50000 T
±35.0000 kOe
±2.80000 MA/m
±3.50000 kG
±350.000 mT
±3.50000 kOe
±280.000 kA/m
±350.000 G
±35.0000 mT
±350.000 Oe
±28.0000 kA/m
±35.0000 G
±3.50000 mT
±35.0000 Oe
±2.80000 kA/m
±3.50000 G
±350.000 µT
±3.50000 Oe
±280.000 A/m
Ultra-High Sensitivity Probe (UHS)
Gauss
Tesla
Oersted
Amp/meter
Range and Resolution
Range and Resolution
Range and Resolution
Range and Resolution
±35.0000 G
±3.50000 mT
±35.0000 Oe
±2.80000 kA/m
±3.50000 G
±350.000 µT
±3.50000 Oe
±280.000 A/m
±350.000 mG
±35.0000 µT
±350.000 mOe
±28.0000 A/m
±35.0000 mG
±3.50000 µT
±35.0000 mOe
±2.80000 A/m
For manual ranging, press the
Select Range
key. The range setup screen appears as a prompt for changing the range.
Sel ec t Wi t h ° ®
Ra n g e + 3 . 5 0 0 0 k G
Use the
s
or
t
key to select from the available ranges. Press Enter to accept the new selection and return to the normal
display. Press Escape to cancel the new selection and return to the normal display.
Содержание 475
Страница 16: ...Lake Shore Model 475 Gaussmeter User s Manual 1 6 Introduction 1 4 SAFETY SYMBOLS ...
Страница 25: ...Lake Shore Model 475 Gaussmeter User s Manual Background 2 9 Figure 2 8 Examples of PI Control ...
Страница 36: ...Lake Shore Model 475 Gaussmeter User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 475 Gaussmeter User s Manual 4 32 Operation This Page Intentionally Left Blank ...
Страница 82: ...Lake Shore Model 475 Gaussmeter User s Manual 6 6 Remote Operation Figure 6 2 Model 475 Status System ...