Lake Shore Model 475 Gaussmeter User’s Manual
Remote Operation
6-27
6.3.1
Interface Commands (Alphabetical Listing)
*CLS
Clear Interface Command
Input:
*
CLS [term]
Remarks:
Clears the bits in the Standard Event Status Register and Operation Event Register and terminates all
pending operations. Clears the interface, but
not
the instrument. The related instrument command is
*
RST
.
*ESE
Standard Event Status Enable Register Command
Input:
*
ESE <bit weighting>[term]
Format:
nnn
Remarks:
The Standard Event Status Enable Register determines which bits in the Standard Event Status
Register will set the summary bit in the Status Byte. This command programs the enable register using
a decimal value which corresponds to the binary-weighted sum of all bits in the register. Refer to
Paragraph 6.1.4.2.1.
*ESE?
Standard Event Status Enable Register Query
Input:
*
ESE? [term]
Returned:
<bit weighting>[term]
Format:
nnn
(Refer to command for description)
*ESR?
Standard Event Status Register Query
Input:
*
ESR? [term]
Returned:
<bit weighting>
Format:
nnn
Remarks:
Bits in this register correspond to various system events and latch when the event occurs. When an
event bit is set, subsequent events corresponding to that bit are ignored. Set bits remain latched until
the register is reset by this query or a *CLS command. Refer to Paragraph 6.1.4.2.1.
*IDN?
Identification Query
Input:
*
IDN? [term]
Returned:
<manufacturer>,<model>,<serial>,<date>[term]
Format:
aaaa,aaaaaaaa,aaaaaaa,mmddyyyy
<manufacture>
Manufacturer ID
<model>
Instrument model number
<serial>
Serial number
<date>
Instrument firmware revision date
Example:
LSCI,MODEL475,1234567,06122003
*OPC
Operation Complete Command
Input:
*
OPC [term]
Remarks:
Used in conjunction with bit 0 (OPC of the Standard Event Status Register. If sent as the last command
in a command sequence, bit 0 will be set when the instrument completes the operation that was
initiated by the command sequence. Refer to Paragraph 6.1.4.3.6 for more information.
Содержание 475
Страница 16: ...Lake Shore Model 475 Gaussmeter User s Manual 1 6 Introduction 1 4 SAFETY SYMBOLS ...
Страница 25: ...Lake Shore Model 475 Gaussmeter User s Manual Background 2 9 Figure 2 8 Examples of PI Control ...
Страница 36: ...Lake Shore Model 475 Gaussmeter User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 475 Gaussmeter User s Manual 4 32 Operation This Page Intentionally Left Blank ...
Страница 82: ...Lake Shore Model 475 Gaussmeter User s Manual 6 6 Remote Operation Figure 6 2 Model 475 Status System ...