Lake Shore Model 475 Gaussmeter User’s Manual
4-26
Operation
To configure the beeper, continue from the alarm inside/outside setup screen or
press and hold
Alarm
and then press
Enter
until the following screen appears as a prompt for configuring the beeper.
Sel ec t Wi t h ° ®
Al a r m Au d i b l e Of f
Use the
s
or
t
key to select from Off or On. Press
Enter
to accept the new selection and continue to the next setting
screen. Press
Escape
to cancel the new selection and return to the normal display.
The Model 475 can be configured to display pass or fail when it is used during repetitive magnet testing or sorting
operations. The sort message can be turned on or off as necessary and does not affect other operations of the alarm
feature. When sort is turned on the lower display line will show:
Fai l Low
Field below low alarm setpoint.
* * Pas s * *
Field between the two alarm setpoints.
Fai l Hi gh
Field above high alarm setpoint.
Refer to Paragraph 4.4.1 to configure the display for the Sort Message.
To enable the sort message, continue from Alarm Audible screen or
press and hold
the
Alarm
and then press
Enter
until
the following screen appears as a prompt for configuring the sort message.
Se l e c t Wi t h ° ®
Al a r m So r t Of f
Use the
s
or
t
key to select from Off or On. Press
Enter
to accept the new selection and return to the normal display.
Press
Escape
to cancel the new selection and return to the normal display.
4.11 RELAYS
The Model 475 has two mechanical relays designated Relay 1 and Relay 2. The relays are most commonly associated
with the high and low alarms, but they can also be controlled manually and used for other purposes. Both relays default
to their normal state when instrument power is turned off.
The relays have two modes of operation:
Manual:
The relay is turned off (normal) or on (active) manually from the front panel or over computer interface.
Alarm:
The relay is configured to follow the operation of the alarms. The user can tie the action of the relay to the
high alarm, low alarm or both alarms. The relay is activated any time the associated alarm is active.
To begin the relay setup process, press the
Relay
key. The following relay setup screen appears as a prompt for relay
selection.
Sel ec t Wi t h ° ®
Re l a y 1
Use the s or t key to select Relay 1 or Relay 2. Press
Enter
to accept the new selection and continue to the next setting
screen. Press
Escape
to cancel the new selection and return to the normal display.
Содержание 475
Страница 16: ...Lake Shore Model 475 Gaussmeter User s Manual 1 6 Introduction 1 4 SAFETY SYMBOLS ...
Страница 25: ...Lake Shore Model 475 Gaussmeter User s Manual Background 2 9 Figure 2 8 Examples of PI Control ...
Страница 36: ...Lake Shore Model 475 Gaussmeter User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 475 Gaussmeter User s Manual 4 32 Operation This Page Intentionally Left Blank ...
Страница 82: ...Lake Shore Model 475 Gaussmeter User s Manual 6 6 Remote Operation Figure 6 2 Model 475 Status System ...