Lake Shore Model 475 Gaussmeter User’s Manual
Service
8-15
13.
Configure Analog Output 2 to manual mode, -50% (MOUT 1,-50).
14.
Read the voltage using the DVM. Verify the reading to be –2.5 VDC, ±1 mV. Record the reading.
15.
Configure Analog Output 2 to manual mode, +50% (MOUT 1,50).
16.
Read the voltage using the DVM. Verify the reading to be +2.5 VDC, ±1 mV. Record the reading.
17.
Configure Analog Output 2 to manual mode, 0% (MOUT 1,0).
18.
Read the voltage using the DVM. Verify the reading to be 0 VDC, ±1 mV. Record the reading.
19.
Send the CALSAVE command to save the calibration constants into non-volatile memory.
8.11.4.2
Analog Output 3 Calibration
1.
Set Analog Output 3 Gain Correction Factor to 1 (CALG 4,3,1) and Offset Correction Factor to 0 (CALZ 4,3,0).
2.
Connect the DVM to Analog Output 3.
3.
Configure Analog Output 3 to manual mode, -100% (ANALOG 3,2,0,0,-100,10).
4.
Read the voltage using the DVM. Record as VMIN.
5.
Configure Analog Output 3 to manual mode, +100% (ANALOG 3,2,0,0,100,10).
6.
Read the voltage using the DVM. Record as VMAX.
7.
Calculate the Gain Correction Factor (GCF) = 20 / (VMAX – VMIN). Note: Gain is calculated over the full span of
the output. Test this factor to be 1.0, ±1%. Record this value.
8.
Send the Gain Correction Factor (GCF) to the Model 475 (CALG, 4,3,<GCF>).
9.
Configure Analog Output 3 to manual mode, 0% (ANALOG 3,2,0,0,0,10).
10.
Read the voltage using the DVM. Record as VZERO.
11.
Calculate the Offset Correction Factor (OCF) = -(VZERO) * (100% / 10). Note: The offset voltage is expressed as a
percentage of Full Scale. Test this factor to be between -1 and 1. Record this value.
12.
Send the Offset Correction Factor (OCF) to the Model 475 (CALZ 4,3,<OCF>).
13.
Configure Analog Output 3 to manual mode, -100% (ANALOG 3,2,0,0,-100,10).
14.
Read the voltage using the DVM. Verify the reading to be –10 VDC, ±1 mV. Record the reading.
15.
Configure Analog Output 3 to manual mode, +100% (ANALOG 3,2,0,0,100,10).
16.
Read the voltage using the DVM. Verify the reading to be +10 VDC, ±1 mV. Record the reading.
17.
Configure Analog Output 3 to manual mode, 0% (ANALOG 3,2,0,0,0,10).
18.
Read the voltage using the DVM. Verify the reading to be 0 VDC, ±1 mV. Record the reading.
19.
Send the CALSAVE command to save the calibration constants into non-volatile memory.
Содержание 475
Страница 16: ...Lake Shore Model 475 Gaussmeter User s Manual 1 6 Introduction 1 4 SAFETY SYMBOLS ...
Страница 25: ...Lake Shore Model 475 Gaussmeter User s Manual Background 2 9 Figure 2 8 Examples of PI Control ...
Страница 36: ...Lake Shore Model 475 Gaussmeter User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 475 Gaussmeter User s Manual 4 32 Operation This Page Intentionally Left Blank ...
Страница 82: ...Lake Shore Model 475 Gaussmeter User s Manual 6 6 Remote Operation Figure 6 2 Model 475 Status System ...