![Lake Shore 475 Скачать руководство пользователя страница 115](http://html.mh-extra.com/html/lake-shore/475/475_user-manual_670635115.webp)
Lake Shore Model 475 Gaussmeter User’s Manual
Remote Operation
6-39
RELAY
Relay Parameter Command
Input:
RELAY <relay number>, <mode>, <alarm type>[term]
Format:
n,n,nn,n
<relay number>
Specifies which relay to configure: 1 = Relay 1, 2 = Relay 2.
<mode>
Specifies relay mode: 0 = Off, 1 = On, 2 = Alarms.
<alarm type>
Specifies the alarm type that activates the relay when the relay is in alarm mode:
1 = Low Alarm, 2 = High Alarm, 3 = Both Alarms.
Example:
RELAY 1,2,2[term]
– Relay 1 is setup in Alarms mode and activates when the low alarm activates.
RELAY?
Relay Parameter Query
Input:
RELAY? <relay number>[term]
Format:
n
<relay number> Specifies which relay to query: 1–2.
Returned:
<mode>, <alarm type>[term]
Format:
n,n
(Refer to command for description)
RELAYST?
Relay Status Query
Input:
RELAYST? <relay number>[term]
Format:
n
<relay number> Specifies which relay to query: 1 = Relay 1, 2 = Relay 2.
Returned:
<status>[term]
Format:
n
0 = Off, 1 = On.
RELSP
Relative Setpoint Command
Input:
RELSP <setpoint>[term]
Format:
±
nnn.nnnE
±
nn
<setpoint> Specifies the setpoint to use in the relative calculation:
±
350 kG
Example:
RELSP 1200[term]
– Configure the relative setpoint as 1200 Gauss(if units in Gauss). The relative
reading will use this value if relative is using the user defined setpoint. Refer to REL command.
RELSP?
Relative Setpoint Query
Input:
RELSP? [term]
Returned:
<setpoint>[term]
Format:
±
nnn.nnnE
±
nn
(Refer to command for description)
TRIG
Hardware Trigger Out Command
Input:
TRIG <off/on>[term]
Format:
n
<off/on>
Specifies hardware trigger off or on. Valid entries: 0 = Off, 1 = on.
Example:
TRIG 1[term]
– The hardware trigger line produces a trigger when the instrument calculates a new
reading.
TRIG?
Hardware Trigger Out Query
Input:
TRIG? [term]
Returned:
<off/on>[term]
Format:
n
(Refer to command for description)
Содержание 475
Страница 16: ...Lake Shore Model 475 Gaussmeter User s Manual 1 6 Introduction 1 4 SAFETY SYMBOLS ...
Страница 25: ...Lake Shore Model 475 Gaussmeter User s Manual Background 2 9 Figure 2 8 Examples of PI Control ...
Страница 36: ...Lake Shore Model 475 Gaussmeter User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 475 Gaussmeter User s Manual 4 32 Operation This Page Intentionally Left Blank ...
Страница 82: ...Lake Shore Model 475 Gaussmeter User s Manual 6 6 Remote Operation Figure 6 2 Model 475 Status System ...