6 OPERATION
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
115
Group
Parameters
Default value
Serial Command
Remote I/O
Remote I/O enabled
Off
N/A
Digital Input 0-7 (8)
See Remote I/O default options
(Mantissa, Exponent)
Digital Input Trigger (8)
Digital Output 0-7 (8)
Digital Output Trigger (8)
Analog Input 0-1 (2)
Analog Output Type
Voltage (Voltage or Current)
Vent
Manual Vent Verify
Off
N/A
AutoTest
Save Results
Off
N/A
Display
Decimal Places
1
N/A
Screen Dim
None (None, 5, 10, 20, 30)
N/A
Show Vacuum Pressure
On
N/A
Option Button
Vent (None, Zero, Vent, Data Log
On, Alarm)
@OP
Language
English
@LA
Leak Detector Settings
Group
Parameters
Default value
Serial Command
Test Method
Test Method
Vacuum (Vacuum, Sniff)
TST
Vacuum Mode
Ultra (Gross, Fine, Ultra)
CYT
Ultra Crossover
0.5 mbar
P2
Fine Crossover
5 mbar
P5
Gross Crossover
25 mbar
P1
Massive Mode
Off
MAS
Reject Limits
Vacuum Setpoint
1.0e-8
S1H
Sniff Setpoint
1.0e-4
S1S
Probe Clogged
1.0e-6
S6
Gross Leak Setpoint
1.0e-5, Off
AA
Gross Leak Auto Cleanup
3.0e-4, On
AP
Vent
Vent Method
Automatic (Manual, Automatic)
IVP
Vent Delay
0
Vent Timer
Off (0 seconds)
AutoTest
Cycle End Method
Manual (Manual, Automatic)
CA
Test Time
10 seconds
Roughing Time
On, 10 seconds
Sniff Auto Test
CAS
Содержание TITAN VERSA C
Страница 1: ...TITAN VERSA LEAK DETECTORS OPERATIONS AND MAINTENANCE MANUAL ...
Страница 10: ......
Страница 28: ...4 DESCRIPTION 18 TITAN VERSA Operations and Maintenance Manual LACO Technologies Vacuum Schematic Main Version ...
Страница 77: ...6 OPERATION LACO Technologies TITAN VERSA Operations and Maintenance Manual 67 Method Settings ...
Страница 82: ...6 OPERATION 72 TITAN VERSA Operations and Maintenance Manual LACO Technologies AutoTest Settings ...