8 MAINTENANCE AND SERVICE
146
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
Figure 14:
TITAN VERSA: Oil Wick Access
a)
O-ring
b)
Operating fluid reservoir
c)
Porex rod
d)
Ventilating screw
10.
Remove the contamination on the turbo molecular pump and cover with a clean,
lint-free cloth. Do not use cleaning liquids.
11.
Insert the new Porex rods (8x) using tweezers.
12.
Slide in the new oil wick cartridge into the pump up to the o-ring.
NOTICE:
Material damage due to incorrect position of the oil wick cartridge.
•
Do not press in the new oil wick cartridge completely. The new oil wick
cartridge will be positioned correctly when you screw in the cover.
13.
Screw in the cover with a new o-ring. Note the tightening torque for the cover of
13 Nm +/-10%.
14.
Make sure that the new o-ring is inserted in the correct position. An incorrectly
mounted o-ring can result in gross leaks and a malfunction of the device.
15.
Reinstall all previous parts.
16.
Reconnect the power supply and power on unit.
17.
Verify proper turbo pump performance.
18.
Enter your maintenance work on the TMP in Maintenance Tasks section.
Содержание TITAN VERSA C
Страница 1: ...TITAN VERSA LEAK DETECTORS OPERATIONS AND MAINTENANCE MANUAL ...
Страница 10: ......
Страница 28: ...4 DESCRIPTION 18 TITAN VERSA Operations and Maintenance Manual LACO Technologies Vacuum Schematic Main Version ...
Страница 77: ...6 OPERATION LACO Technologies TITAN VERSA Operations and Maintenance Manual 67 Method Settings ...
Страница 82: ...6 OPERATION 72 TITAN VERSA Operations and Maintenance Manual LACO Technologies AutoTest Settings ...