5 INSTALLATION
30
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
5.2.5.2.
Pump Exhaust
WARNING:
Danger from exhaust gases and vapors
•
Exhaust gases and vapors from oil-sealed pumps may be harmful to health.
•
When operating the device in poorly ventilated spaces, mount an exhaust gas
line on the TITAN VERSA.
The pump exhaust for VERSA wet pump versions consists of a 12 mm push tube
connection. If the VERSA is operated in poorly ventilated spaces, an exhaust gas line is
recommended. An exhaust gas line is not included in the scope of delivery. It is the duty
of the owner to provide an exhaust gas line and to ensure the discharge of exhaust
gases. Do not connect long tube sections to this port; this can adversely affect pump
performance. Use short, large-diameter tubing sections (see P/N: LMSA5503 for 6-
foot section).
Dry pump versions contain a muffler for the pump exhaust and are to be left alone.
Mufflers should be replaced every two years under typical operations.