6 OPERATION
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
75
Background Suppression
The background suppression feature subtracts the background reading using the value
from the end of the last successful calibration. This function works both in Standby and
Test modes. This feature is enabled by default and can be used to measure a leak rate
two decades lower than the detector's background.
Data Log Settings
NOTICE:
USB drive
For data logging features to function, a USB drive must be inserted in the USB
port. A USB is provided in shipment in documentation packet. Spare drives can
be ordered (P/N: TV5928).
Содержание TITAN VERSA C
Страница 1: ...TITAN VERSA LEAK DETECTORS OPERATIONS AND MAINTENANCE MANUAL ...
Страница 10: ......
Страница 28: ...4 DESCRIPTION 18 TITAN VERSA Operations and Maintenance Manual LACO Technologies Vacuum Schematic Main Version ...
Страница 77: ...6 OPERATION LACO Technologies TITAN VERSA Operations and Maintenance Manual 67 Method Settings ...
Страница 82: ...6 OPERATION 72 TITAN VERSA Operations and Maintenance Manual LACO Technologies AutoTest Settings ...