5 INSTALLATION
24
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
VERSA C External Pump Connections
The VERSA C (Compact) configuration does not include a primary vacuum pump.
The user must supply the following items in this configuration:
•
Primary pump meeting the following specifications:
▪
Ultimate pressure below 3 mbar
▪
Pumping speed greater than 2 m3/h
•
Vacuum hose with KF-25 end connection
•
Power connection to external primary pump