8 MAINTENANCE AND SERVICE
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
147
Replace SplitFlow 50 for Bearing Change
Bearing changes are recommended at least every four years. Under severe usage this
may need to be done more frequently. A bearing change can only be carried out by
LACO Service Center.
CAUTION:
Ensure all maintenance is performance properly per
instructions in this manual.
Check Oil Level and Color
Check the pump oil level and color monthly, or more frequently under heavy usage.
Change the oil if the level is below the minimum level mark. Oil level verification is
most accurate if checked while pump is off, hot, and on a horizontal plane.
LACO recommends changing the pump oil if the oil color appears to be heavily
discolored or white in appearance.
Accessing Uno6 pump
Most pump maintenance requires the covers to be removed. The Uno6 pump is in the
bottom section of the leak detector. With the covers off, LACO recommends cleaning
the interior area and pump surfaces.
8.16.2.1.
Removing Pump
1.
Place leak detector on blocks to gain access to pump bolts under the unit.
2.
Remove lower tower covers.