8 MAINTENANCE AND SERVICE
128
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
LACO Advanced Service Options
Model/Pump
PM Services P/N
Service Agreements
VERSA L/wet pump
LS-TVL-3 (2-3 year)
LS-TVL-4 (3 year)
LSA-TVL-2 (2 year)
LSA-TVL-3 (3 year)
LSA-TVL-4 (4 year)
VERSA L/dry pump
LS-TVLD-3 (2-3 year)
LS-TVLD-4 (3 year)
LSA-TVLD-2 (2 year)
LSA-TVLD-4 (4 year)
VERSA T/wet pump
LS-TVT-3 (2-3 year)
LS-TVT-4 (3 year)
LSA-TVT-2 (2 year)
LSA-TVT-3 (3 year)
LSA-TVT-4 (4 year)
VERSA T/dry pump
LS-TVTD-3 (2-3 year)
LS-TVTD-4 (3 year)
LSA-TVTD-2 (2 year)
LSA-TVTD-4 (4 year)
Front Cover (all Models)
7.
To remove front cover, you must first remove the back cover (see Sections 8.8.2
and 8.8.3 below).
8.
Remove the two bolts on either side of the front cover with a 5 mm Hex wrench.
9.
Remove the test port connections and rubber dust cover.
10.
Lift cover until the top is above the test port, then lift at a forward angle.
11.
Carefully pull cover away from leak detector to locate and detach the 40 pin IDC
screen cable from screen.