4 DESCRIPTION
16
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
Technical Data
Parameter
VERSA C
VERSA T
VERSA TD
VERSA L
VERSA LD
Primary pump type
None
Wet
Dry
Wet
Dry
Primary pump speed m
3
/h (cfm)
N/A
6.0 (3.5)
3.6 (2.1)
15 (8.8)
6.5 (3.8)
Start-up time
~ 3 min (@ 20 °C)
Detectable gases
Helium 4, Helium 3, Hydrogen
Test methods
Vacuum and sniff
Helium pumping speed (L/s)
2.5
Minimum detectable leak rate in
vacuum, helium (mbar L/sec)
5 x 10
-12
Minimum detectable leak rate in
sniff, helium (mbar L/sec)
5 x 10
-9
Vacuum test modes with max inlet
test pressure
Gross
–
25 mbar
Fine
–
5 mbar
Ultra
–
0.5 mbar
Protection category
IP 20
Inlet flange
DN 25 ISO-KF
Filament
2 (iridium yttriated)
Turbo pump flow
–
SplitFlow 50
53 L/s N
2
Operating temperature range -
Vacuum (°C)
0 - 45
0 - 45
0 - 45
0 - 45
0 - 45
Operating temperature range -
sniff (°C)
0 - 45
0 - 40
0 - 40
0 - 45
0 - 45
Noise level (dB A)
N/A
58
56
57
58
Electrical supply
–
V~, 50/60 Hz
90 - 250
100 - 130
208 - 240
90-250
100 - 130
208 - 240
100
–
130
208 - 240
Electrical power max (W)
350
700
600
850
600
Weight kg (lbs)
25 (55)
44 (97)
42 (93)
58 (128)
47 (104)
Dimensions mm (H x W X D) (in)
403 x 562 x 323
(15.8 x 22.1 x 12.7)
607 x 562 x 323
(23.9 x 22.1 x 12.7)
403 x 562 x 497
(15.8 x 22.1 x 19.6)
Содержание TITAN VERSA C
Страница 1: ...TITAN VERSA LEAK DETECTORS OPERATIONS AND MAINTENANCE MANUAL ...
Страница 10: ......
Страница 28: ...4 DESCRIPTION 18 TITAN VERSA Operations and Maintenance Manual LACO Technologies Vacuum Schematic Main Version ...
Страница 77: ...6 OPERATION LACO Technologies TITAN VERSA Operations and Maintenance Manual 67 Method Settings ...
Страница 82: ...6 OPERATION 72 TITAN VERSA Operations and Maintenance Manual LACO Technologies AutoTest Settings ...