6 OPERATION
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
95
•
U = 6.9 V -> Mantissa = 6.9
Exponent
The "Exponent" output corresponds with the leak rate exponent:
•
it increases by 1 V per decade,
•
the starting decade corresponds with 10 V.
Formula
U = Voltage measured (V) on analog output D
0
= Low decade for 0 V
Exponent = 10
–
U + D
0
Example 1
Low decade at 10-12 (10 V = -12) -> D
0
= -12
•
U = 7 V -> Exponent = 10 - 7 - 12 -> Exponent = -9
•
U = 2 V -> Exponent = 10 - 2 - 12 -> Exponent = -4
Example 2
Low decade at 10-10 (10 V = -10) -> D0 = -10
•
U = 7 V -> Exponent = 10 - 7 - 10 -> Exponent = -7
•
U = 2 V -> Exponent = 10 - 2 - 10 -> Exponent = -2
Logarithmic
The "Logarithmic" output corresponds with the leak rate value:
•
it increases by 1 V per decade,
•
the starting decade corresponds with 0 V.
Formulas
U = Voltage measured (V) on analog output D0 = Low decade for 0 V
Mantissa = 10
(U- Integer value (U))
Exponent = Integer value (U) + D
0
Leak rate = Mantissa x 10
Exponent
Example 1
Low decade at 10-12 (0 V = 1 10-12) -> D0 = -12
•
V = 3 91 V -> Leak rate = 10
(3 91-3)
x 10
(3-12 )
= 8 13 10
-9
•
V = 8 25 V -> Leak rate = 10
(8 25-8)
x 10
(8-12 )
= 1 78 10
-4
Example 2
Low decade at 10-10 (0 V = 1 10-10) -> D0 = -10
•
V = 3 91 V -> Leak rate = 10(3 91-3) x 10
(3-10 )
= 8 13 10
-7
•
V = 8 25 V -> Leak rate = 10(8 25-8) x 10
(8-10 )
= 1 78 10
-2
Compound
The "He Compound" output is a combination of mantissa and exponent:
•
the integer part represents the exponent
•
the decimal part represents the mantissa.
Formulas
U = Voltage measured (V) on analog output
Mantissa = 10 x (U - Integer value (U))
Содержание TITAN VERSA C
Страница 1: ...TITAN VERSA LEAK DETECTORS OPERATIONS AND MAINTENANCE MANUAL ...
Страница 10: ......
Страница 28: ...4 DESCRIPTION 18 TITAN VERSA Operations and Maintenance Manual LACO Technologies Vacuum Schematic Main Version ...
Страница 77: ...6 OPERATION LACO Technologies TITAN VERSA Operations and Maintenance Manual 67 Method Settings ...
Страница 82: ...6 OPERATION 72 TITAN VERSA Operations and Maintenance Manual LACO Technologies AutoTest Settings ...