31
Manual NO. 220412L002 Version : 0.0
(6) Rotate the control knob to the end and select ESC (or no input for 20seconds) to go back to
main display.
If the error occurs such as exceed the range of over temperature during temperature control,
the control stops.
As the unit calibrate the temperature in many points, if you input wrong adjustment, the unit
can’t proceed the calibration.
You can see the result of the calibration any time in CALIB VIEW. It displays C1, C2, C3for
checking measure points. POINT represents measure points, VALUE represents the setting
value of calibration.
RESET CALIB can reset each Calibration
4.6.3 Temp Deviation
The unit alarms for Deviation warning when the difference between SV and PV value is over
allowable range. After the alarm, although the temperature restored to be normal, it takes time for end
of Deviation alarm.
Содержание IL3-15
Страница 1: ...OPERATION MAUNAL IL3 series Model IL3 15 IL3 25 Manual No 220412L002 Version 0 0 ...
Страница 6: ......
Страница 7: ...1 Manual NO 220412L002 Version 0 0 1 0 Safety ...
Страница 17: ...11 Manual NO 220412L002 Version 0 0 2 0 Functional Description ...
Страница 19: ...13 Manual NO 220412L002 Version 0 0 2 3 Structure IL3 15 IL3 25 ...
Страница 21: ...15 Manual NO 220412L002 Version 0 0 3 0 Installation ...
Страница 26: ...20 2016 All Rights Reserved Jeio Tech Co Ltd 4 0 Operation ...
Страница 43: ...37 Manual NO 220412L002 Version 0 0 5 0 Maintenance ...
Страница 46: ...40 2016 All Rights Reserved Jeio Tech Co Ltd 6 0 Troubleshooting ...
Страница 49: ...43 Manual NO 220412L002 Version 0 0 7 0 Jeiotech Software ...
Страница 52: ...46 2016 All Rights Reserved Jeio Tech Co Ltd 8 0 Appendix ...