User can check the parameter setting value 9 using (Config button), and also change the parameter
setting value (by Up/Down button)
Lower limit: -10 ~ upper limit (by Celsius)
Upper limit: lower limit to 120 degrees (Celsius criteria)
The screen will be returned to the main screen, If the product is standby condition for 5 sec.
If the temperature limiter shown value is different with actual value (measured calibration temperature
device),
Please corrected in the following way;
a. Shut if the power of the product.
b. Press the Config button with Up/Down button, and turn on the main power.
c. About 4 sec later, every setting will be initialization and show OFFSET setup screen.
d. User can set OFFSET value by Config button.
e. Set the OFFSET value (-10~10
Celsius criteria
) by using Up/Down button
f. Save the Config value.
g. Supply the main power again to product, the product will be applied by changed OFFSET value.
4.8.2 Current Protection
Fuse : it can automatically shut off the current when overcurrent condition immediately.
4.8.3 Door Switch
The unit heater and fan will be stopping when door is open automatically.
Содержание IL3-15
Страница 1: ...OPERATION MAUNAL IL3 series Model IL3 15 IL3 25 Manual No 220412L002 Version 0 0 ...
Страница 6: ......
Страница 7: ...1 Manual NO 220412L002 Version 0 0 1 0 Safety ...
Страница 17: ...11 Manual NO 220412L002 Version 0 0 2 0 Functional Description ...
Страница 19: ...13 Manual NO 220412L002 Version 0 0 2 3 Structure IL3 15 IL3 25 ...
Страница 21: ...15 Manual NO 220412L002 Version 0 0 3 0 Installation ...
Страница 26: ...20 2016 All Rights Reserved Jeio Tech Co Ltd 4 0 Operation ...
Страница 43: ...37 Manual NO 220412L002 Version 0 0 5 0 Maintenance ...
Страница 46: ...40 2016 All Rights Reserved Jeio Tech Co Ltd 6 0 Troubleshooting ...
Страница 49: ...43 Manual NO 220412L002 Version 0 0 7 0 Jeiotech Software ...
Страница 52: ...46 2016 All Rights Reserved Jeio Tech Co Ltd 8 0 Appendix ...