12
©2016. All Rights Reserved. Jeio Tech Co., Ltd
2.1 Introductions
This equipment is a multi-purpose low temperature incubator for biotechnology, pharmacy, medical
science, chemistry, and biology. It has applied CLS (Custom Logical Safe) – Control System that Jeio
Tech originally developed for the convenience in use and the s
2.2 Features
1) CLS-Control system
1
means “Control system which has logical safety device specialized for
individual model”. Laboratory must have Thermal safety secure because there are a lot of
inflammable reagent. This system is highest safety secure control device and makes the unit
suitable for this kind of environment.
(2) CLS-Control System carries out temperature control of equipment and heating control in the Main
CPU where precise PID algorithm is possible. All actions for safety are conducted by a selective
functional Logic IC which is installed separately. This is designed to conduct safety performance
against any electric and electronic shock on the unit.
(3) CLS-Control System shuts down all 2 phase power supply to each part immediately and informs
user instability by audible and visual device, and then it keeps in safety mode until all instability
conditions are removed.
(4) CLS-Control System gives user two choices when the unit’s operation is terminated by a power
failure and then the power recovers. One is a resume of the unit’s operation. Another one is to keep
the unit in standstill.
(5) CLS-Control System is composed that feeble electric current streams down to only 5V, 10mA at a
point of contact of a used safety device. There is no damage of a point of contact in use of a long
time. So, the durability is very long. After it stops the control of the Thyristor which controls electric
current stream of the heater that uses a large electric current when a safety device is operated, it
shuts down the power through Magnet Switch’s disconnection of a point of contact. It prevents the
point of contact from damaging and a noise revelation of a power switch. It enforced with the
mentioned objection order above when an operation of equipment starts, and it also prevents the
point of contact from damaging and a noise revelation of Magnet Switch.
1
Control system which has logical safety device specialized for individual model
Содержание IL3-15
Страница 1: ...OPERATION MAUNAL IL3 series Model IL3 15 IL3 25 Manual No 220412L002 Version 0 0 ...
Страница 6: ......
Страница 7: ...1 Manual NO 220412L002 Version 0 0 1 0 Safety ...
Страница 17: ...11 Manual NO 220412L002 Version 0 0 2 0 Functional Description ...
Страница 19: ...13 Manual NO 220412L002 Version 0 0 2 3 Structure IL3 15 IL3 25 ...
Страница 21: ...15 Manual NO 220412L002 Version 0 0 3 0 Installation ...
Страница 26: ...20 2016 All Rights Reserved Jeio Tech Co Ltd 4 0 Operation ...
Страница 43: ...37 Manual NO 220412L002 Version 0 0 5 0 Maintenance ...
Страница 46: ...40 2016 All Rights Reserved Jeio Tech Co Ltd 6 0 Troubleshooting ...
Страница 49: ...43 Manual NO 220412L002 Version 0 0 7 0 Jeiotech Software ...
Страница 52: ...46 2016 All Rights Reserved Jeio Tech Co Ltd 8 0 Appendix ...