3 Operation Method - Full-Auto Protocol
56
QuickGene-Auto240L Operation Manual
3
(2)
Select the sample suction starting position.
Select [BOTTOM] for sample suction from the bottom of
the blood collection tube; select [SURFACE] for suction
from the fluid surface.
[BOTTOM]: Suction from a fixed position
[SURFACE]: Suction from fluid surface by detecting the
surface using the differential pressure type fluid
surface detecting function
Note: When [BOTTOM] mode is selected, a pop-up
window to warn of overflow will appear. After
confirming that the sample quantity in the blood
collection tube is not more than 3 mL, press [OK]
and return to the setup screen.
If the set sample quantity is more than 3 mL, press
[BACK] to go back to the mode select screen, press
[UNLOCK] to release the door locks, and then take out
the sample holder from the system and adjust the
sample quantity.