3 Operation Method - Full-Auto Protocol
50
QuickGene-Auto240L Operation Manual
3
<2> Enter Sample Information
to A-1
to B-1
A. When Sample ID reading function is ON
(A-1)
When the door lock is released, open the right flap door.
(A-2)
Insert sample holder A in slot A slowly over approx. 8
seconds. The ID is read.
Securely set until it contacts the
stopper on the end.
Do not peer into the system during sample ID reading and
view the red LED light source for the barcode reader.
Direct
viewing at length may cause eye disorders.
Note:
・
When setting a holder in the slot, securely set until it
contacts the stopper on the end
.
・
When several holders are used, pay attention to the
holder identification symbols A-C for setting.
(A-1)
The reading of complete samples will be displayed in
reversed deep blue on the operation screen.
Check for the correctness of readout positional
information and the position of the set sample.
(A-4)
If there are more holders to set, press
[NEXT]
and carry out
(A-2), (A-3)
for holders
B, C.
Sample ID Read
Function ON
Sample ID Read
Function OFF
Warning
Barcode Reader (Inside of system)
Right flap door
Содержание QuickGene-Auto240L
Страница 1: ...Automated Nucleic Acid Isolation System QuickGene Auto240L Ver 1 0 1 Operation Manual ...
Страница 2: ......
Страница 14: ...xii QuickGene Auto240L Operation Manual ...
Страница 44: ...2 Basic Functions 30 QuickGene Auto240L Operation Manual 2 ...
Страница 78: ...3 Operation Method Full Auto Protocol 64 QuickGene Auto240L Operation Manual 3 ...