3 Operation Method - Full-Auto Protocol
52
QuickGene-Auto240L Operation Manual
3
<3> Enter Collection Information
to C-1
to D-1
C. When the collection ID reading function is ON
(C-1)
Open the left flap door
(C-2)
Insert the collection tube holder A in the collection ID
reading slot slowly over approx. 8 seconds. The ID is
read.
・
Securely set until it contacts the stopper on the end.
・
Position of the collection tubes completed with reading
will be displayed in reversed green on the operation
screen.
・
When the position of the read-out collection tube and the
previously entered sample setting position.
The laser light source categorized Class 1 is used for the
2D code reader. Do not peer into the system while
reading the collection ID and directly view the red light
source. Direct viewing at length may cause eye
disorders.
Note: If a pop-up window is not displayed, check the set
position of collection tubes and execute the reading
again.
Collection ID Read
Function ON
Collection ID Read
Function OFF
Left flap door
Warning
:
2D Barcode Reader