3 Operation Procedures - Full Auto Protocols
55
QuickGene-Auto240L Operation Manual
3
<5> Setting up Operating Conditions
(1)
Setup of sample tip use starting position.
Press [
RESET]
to start from the default position, and then
set the value to “1”. To start from another position, press
[CHANGE]
to display a ten-key pad, enter a chosen value
and press
[ENT]
to set up.
Refer to the figure on the left for the relation of set-up
values and tip positions.
The number of used sample tips is recorded based on the
operation history; the use starting position will be
automatically set up accordingly.
Note:
・
Confirm whether the number of 1.2-mL tips is equal
to or more than the number of samples and set from
the use starting position.
・
If the sum of setup value and the number of samples
automatically operated exceeds 96, reset the value
of use starting position.
81
65
49
33
17
1
82
66
50
34
18
2
83
67
51
35
19
3
84
68
52
36
20
4
85
69
53
37
21
5
86
70
54
38
22
6
87
71
55
39
23
7
88
72
56
40
24
8
89
73
57
41
25
9
90
74
58
42
26
10
91
75
59
43
27
11
92
76
60
44
28
12
93
77
61
45
29
13
94
78
62
46
30
14
95
79
63
47
31
15
96
80
64
48
32
16
Relation of Setup Values and Tip Positions
Содержание QuickGene-Auto240L
Страница 1: ...Automated Nucleic Acid Isolation System QuickGene Auto240L Ver 1 0 1 Operation Manual ...
Страница 2: ......
Страница 14: ...xii QuickGene Auto240L Operation Manual ...
Страница 44: ...2 Basic Functions 30 QuickGene Auto240L Operation Manual 2 ...
Страница 78: ...3 Operation Method Full Auto Protocol 64 QuickGene Auto240L Operation Manual 3 ...