![Kurabo QuickGene-Auto240L Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/kurabo/quickgene-auto240l/quickgene-auto240l_operation-manual_2011195016.webp)
1 System Installation Procedures
2
QuickGene-Auto240L Operation Manual
1
A location with a secure power source (This system is an overvoltage category II device. Be sure to use the
included power cable.)
・
Operating temperature 15°C-30°C, room humidity 30%-80% (no condensation)
・
A stable horizontal location with load-bearing capacity of
≧
350kg and minimal vibration.
・
A location not exposed to direct sunlight or strong light (Use a curtain, blind, etc. for obstructing
light as necessary.)
・
A location with good ventilation and minimal dust
・
A location with no abrupt variation in temperature (Swift heating of a very cold room or moving
this system from a lower temperature to a warm place, etc. will generate water droplets in the
system (condensation), and the isolation result will be adversely influenced.)
・
A location where temperature and humidity within the specified ranges can be maintained (Water
faucets, boilers, coolers, heaters, etc. should not be in proximity.)
・
A location where strong magnetism (motors, transformers, televisions, speakers, magnets, etc.) is
not in proximity (Approaching a magnetic source will cause malfunction.)
■ Operating Environment
Item
Specification
Temperature (
o
C)
In operation
15-30
During standby
15-30
Humidity (%)
In operation
30-80
During standby
10-80
Max. Wet-bulb Temp.
(
o
C)
In operation
29 (no condensation)
During standby
29 (no condensation)
Temperature Gradient
(
o
C / Hr)
≦
12 (no condensation)
Humidity Gradient
(% / day)
≦
30 (no condensation)
Altitude (m)
Use at
≦
1600
Содержание QuickGene-Auto240L
Страница 1: ...Automated Nucleic Acid Isolation System QuickGene Auto240L Ver 1 0 1 Operation Manual ...
Страница 2: ......
Страница 14: ...xii QuickGene Auto240L Operation Manual ...
Страница 44: ...2 Basic Functions 30 QuickGene Auto240L Operation Manual 2 ...
Страница 78: ...3 Operation Method Full Auto Protocol 64 QuickGene Auto240L Operation Manual 3 ...