3 Operation Procedures - Full Auto Protocols
51
QuickGene-Auto240L Operation Manual
3
(A-5)
After confirming the reading of all sample ID information,
press [
COMPLETE
].
(A-6)
The number of samples whose
IDs
are read out will be
displayed in a pop-up window. If it is correct, close all
doors and press
[OK]
.
Note:
Do not take out the sample holder until the automatic
operation is complete after this operation.
Premature removal will cause an error, the sample ID
information will be deleted, and the system will go
back to "<2> Enter Sample Information"
(A-7)
Go to the next section "<3> Enter Collection Information "
B. When Sample ID Reading Function is OFF
Note:
When the sample ID reading function is OFF, refer to "3.5 Preparation of Samples" and set
the sample holders before turning the system power ON.
(B-1)
Press the numeric button corresponding to the number
of set samples.
・
The pressed buttons will be indicated in reversed deep
blue on the operation screen.
・
The selection will be canceled if the pressed button is
pressed again.
・
Pressing A, B or C button will enable selection of all for
each holder row.
(B-2)
Check that the selected button coincides with the
number of set samples and press [OK].
(B-3)
Move on to the next section "<3> Enter Collection
Information"
Содержание QuickGene-Auto240L
Страница 1: ...Automated Nucleic Acid Isolation System QuickGene Auto240L Ver 1 0 1 Operation Manual ...
Страница 2: ......
Страница 14: ...xii QuickGene Auto240L Operation Manual ...
Страница 44: ...2 Basic Functions 30 QuickGene Auto240L Operation Manual 2 ...
Страница 78: ...3 Operation Method Full Auto Protocol 64 QuickGene Auto240L Operation Manual 3 ...