Multicheck6010
EN
- 92 -
Principle: ρ= 2•Π•d•Re, where Re is a measured resistance in 4-wire method and d is distance
between the probes.
Rh and Rs values are indicative.
6.9 Datos generales
Power supply voltage
9 VDC (6x1.5 V battery cells, size AA)
Power supply adapter
12 V CC / 1000 mA
Battery charging current
< 600 mA (internally regulated)
Voltage of charged batteries
9 VCC (6x1.5 V, at fully charged state)
Charging duration time
typical 6h
Operation
typical 15 h
Overvoltage category
CAT III / 600 V; CAT IV / 300 V
Protection classification
double insulation
Pollution degree
2
Protection degree
IP 42
Display
TFT LCD de 480X320
COM-Port
USB
Dimensions (w x h x d)
25 cm x 10.7 cm x 13.5 cm
Weight (without battery)
1.30 kg
Reference conditions
Reference temperature range
10 ºC - 30 ºC
Reference humidity range
40 %HR - 70 %HR
Operating conditions
Working temperature range
0 ºC - 40 ºC
Maximum relative humidity
95 %HR (0 ºC - 40 ºC), non-condensing
Storage conditions
Temperature range
-10 ºC - +70 ºC
Maximum relative humidity
90 %HR (-10 ºC - +40 ºC)
80 %HR (40 ºC - 60 ºC)
The error in operating conditions could be at most the error for reference conditions (specified in
the manual for each function) + 1 % of measured value + 1 digit unless otherwise specified.