6.2 Continuity resistance
6.2.1 Low R
Measuring range according to EN61557-4 is 0.16 Ω
÷
1999 Ω.
Range (Ω)
Resolution (Ω)
Accuracy
0.1
÷
20.0
(0.10 Ω ... 19.99 Ω) 0.01 Ω
±(3 % of r 3 digits)
20.0
÷
1999
(20.0 Ω ... 99.9 Ω) 0.1 Ω
(100 Ω ... 1999 Ω) 1 Ω
±(5 % of reading)
Open-circuit voltage
5 VCC
Measuring current
min. 200 mA into load resistance of Ω
Test lead compensation
up to 5 Ω
Number of possible tests
with a new set of batteries
up to 1400 (with 2300mAh battery cells)
Automatic polarity reversal of the test voltage.
6.2.2 Low current continuity
Range (Ω)
Resolution (Ω)
Accuracy
0.1
÷
1999
(0.1 Ω ... 99.9 Ω) 0.1 Ω
(100.0 Ω ... 1999 Ω) 1 Ω
±(5 % of r 3 digits)
Open-circuit voltage
5 VCC
Corriente de cortocircuito
max. 7 mA
Test lead compensation
up to 5 Ω
6.3 RCD testing
6.3.1 General data
Nominal residual current
10 mA, 30 mA, 100 mA, 300 mA, 500 mA, 650mA, 1000 mA
Nominal residual current accuracy -0 / +0.1·IΔ; IΔ = IΔN, 2xIΔN, 5xIΔN
-0.1·IΔ / +0; IΔ = ½xIΔN
Test current shape
Sinusoidal (AC), CC (B), impulso (A)
RCD type
general (G), selective (S, time-delayed)
Test current starting polarity
0º ó 180º
Voltage range
93V-134V; 185V-266V; 45Hz-65Hz