Multicheck6010
EN
- 64 -
disconnected and all switches closed!
•
Do not touch the test object during the measurement or before it is fully discharged! Risk of
electric shock!
•
In order to prevent damaging the test instrument, do not connect test terminals to an external
voltage higher than 550 V (AC or DC).
4.2 Continuity
Two continuity sub-functions are available:
•
R Low, ca. 240mA continuity test with automatic polarity reversal.
•
Low current (ca. 4mA) continuous continuity test, useful when testing inductive systems.
4.2.1 R low test
How to perform a R Low resistance measurement
Step 1
Select the Continuity function with the FCT key and select the R Low mode with the
▲▼
and
◀▶
navigation keys. The following menu will be displayed:
Figure 4 4: R Low resistance measurement menu
Step 2
Ajuste el siguiente valor límite:
•
Límit:
limit resistance value using the
▲▼
and
◀▶
navigation keys.
Step 3
Connect test cable to MUlticheck6010. Before performing an R Low resistance
measurement, compensate for the test leads resistance as follows:
1. Short test leads first as shown in figure 4.5.
Figure 4 5: Shorted test leads
2. Press the COM key. After performing test leads compensation the compensated test
leads indicator
COMP
will be displayed in the status line.
3. In order to remove any test lead resistance compensation, just press the COM key
again. After removing any test lead compensation, the compensation indicator will
disappear from the status line.
Содержание Multicheck6010
Страница 1: ...Multicheck6010 Comprobador multifunción Multifunction tester Manual de usuario User s manual ...
Страница 2: ......
Страница 51: ...Multicheck6010 EN Multifunction tester ...
Страница 99: ......
Страница 100: ......