SensoGate WA131MH
23
5 Operation
5.1 Moving into the PROCESS Position
Note:
When the PROCESS position is reached, this will be indicated in different ways, depending on
the SensoGate WA131MH version.
Note:
The safety lock button pops out when the PROCESS position is reached (see detail B). Only if the
safety lock button has popped out is the function of the safeguard “Immersion lock without a
mounted sensor” ensured.
SERVICE
PROCESS
1
2
A
B
01.
Mount the sensor.
Installing and Removing a Sensor, p. 25
Note:
When the rotary movement starts, the safety lock button is automatically depressed.
02.
Depress the safety lock button
(2)
(see detail A) and rotate the rotating collar
(1)
clockwise.
✓
The safety lock button
(2)
pops out when the PROCESS position is reached (see detail B).
✓
Rotating collar
(1)
is locked to prevent rotation.