6.3.6 Calibration Chamber: Removal
Note:
ZU0754 service set or ZU0740 service set are required to remove the calibration chamber.
1
1
2
3
01.
Remove the process unit from the drive unit.
Drive Unit: Disassembly, p. 31
02.
Remove the screws
(1)
with a screwdriver of type TX25. Keep the screws
(1)
in a safe place for as-
sembly later on.
03.
Position the pliers
(3)
and use the face pin spanner wrench
(2)
to loosen the coupling of the split
calibration chamber.
4
5
04.
Unscrew the top
(4)
from the bottom
(5)
of the calibration chamber and separate the two parts.