SensoGate WA131MH
21
3.3 Outlet Hose: Installation
Note:
The outlet is used to drain trapped process medium and must not be closed. By moving the sen-
sor to the respective limit positions, pressurized process medium can enter the calibration chamber
and be compressed when the outlet is closed. This process medium may splash out during sensor re-
placement.
2
3
4
3
4
1
Note:
The symbol
(1)
designates the outlet.
01.
Push the hose clamp
(3)
onto the outlet hose
(4)
.
02.
Completely push the outlet hose
(4)
onto the connection nipple
(2)
.
03.
Secure the outlet hose
(4)
with the hose clamp
(3)
.
3.4 Inlet Hose: Installation
NOTICE!
Drinking water may be contaminated by rinse and process media when connecting to drink-
ing water pipes.
Observe the information contained in EN 1717. Install a suitable check valve
(e.g., check valve RV01) at the water or rinse connection.
1
2
5
4
3
1
5
4
3
Note:
To ensure safe operation, the sealing insert or the inlet hose
1)
must be installed on the inlet. As
delivered, the inlet port is sealed with a sealing insert.
01.
To install the inlet hose
(5)
, unscrew the sealing insert
(2)
from the inlet port
(1)
(A/F 10 mm).
02.
Screw the coupling
(3)
, part of the inlet hose
(5)
, into the inlet port
(1)
.
03.
Fasten the inlet hose
(5)
with coupling nut
(4)
to the coupling
(3)
.
1)
Availability is dependent on the ordered version