SensoGate WA131MH
31
6.3 Corrective Maintenance
6.3.1 Corrective Maintenance Safety Instructions
WARNING! Process medium, possibly containing hazardous substances, can escape from the
SensoGate WA131MH.
Follow the safety instructions.
CAUTION! Risk of cutting injuries from broken sensor glass.
Handle the sensor with care. Follow
the safety instructions in the related sensor documentation.
6.3.2 Drive Unit: Disassembly
1
2
3
4
5
01.
Safely disconnect the SensoGate WA131MH from the process.
Retractable Fitting: Removal, p. 39
02.
As necessary, disconnect the outlet hose, inlet hose
1)
, and limit switch
1)
.
03.
Move the SensoGate WA131MH into the SERVICE position.
Moving into the SERVICE Position, p. 24
04.
As necessary, remove the sensor
(3)
.
Installing and Removing a Sensor, p. 25
Note:
Do not tilt the coupling nut. Use a suitable spanning wrench (e.g., the one contained in
ZU0680 service set or ZU0740 service set).
05.
Using the spanning wrench
(1)
, loosen the coupling nut
(2)
counterclockwise.
06.
Pull the drive unit
(4)
out of the process unit
(5)
.
1)
Availability is dependent on the ordered version.