6.3.3 Drive Unit: Assembly
Note:
The radial installation position of the drive unit is determined by a coding pin in the calibration
chamber and a hole in the drive unit. The coupling nut can be tightened only if the drive unit is cor-
rectly inserted into the process unit.
1
2
3
4
5
6
7
8
9
01.
Move the drive unit to the SERVICE position.
Moving into the SERVICE Position, p. 24
02.
Push the drive unit
(1)
with the immersion tube
(3)
into the process unit
(5)
. While doing so, posi-
tion the coding pin
(4)
in the hole
(2)
.
Note:
Do not tilt the coupling nut. Use a suitable spanning wrench (e.g., the one contained in
ZU0680 service set or ZU0740 service set).
03.
Position the coupling nut
(8)
and tighten clockwise finger tight or to 10 Nm using the spanning
wrench
(6)
.
04.
As required, install the outlet hose at the outlet
(7)
.
Outlet Hose: Installation, p. 21
05.
Optional: Install the inlet hose
1)
.
Inlet Hose: Installation, p. 21
06.
Optional: Install the limit switch
1)
.
07.
As required, install the sensor
(9)
.
Installing and Removing a Sensor, p. 25
1)
Availability is dependent on the ordered version.