Calibration / Adjustment
Temp Probe Adjustment
Note: With Protos II 4400(X) in the Calibration menu,
with Protos 3400(X) in the Maintenance menu.
Temp Probe Adjustment
This function allows compensating for the individual temperature probe tol-
erance and the influence of the lead resistances to increase the accuracy of
temperature measurement. Make sure that the process temperature is precisely
measured using a calibrated reference thermometer when performing an
adjustment. The measurement error of the reference thermometer should be
less than 0.1 °C. Adjustment without precise measurement might result in con-
siderable deviations of the measured value display!