56
Parameter Setting
ORP/rH value, delta function
Note: Function check (HOLD) mode active
Menu Display
ORP/rH value, delta function
(selection Pg 53)
ORP/rH value
• Select type of reference electrode:
Ag/AgCl, KCl 1 mol/l
(silver/silver chloride)
Ag/AgCl, KCl 3 mol/l
(silver/silver chloride)
Hg, Tl/TlCl, KCl 3.3 mol/l
(Thalamid)
Hg/Hg2SO4, K2SO4 saturated (mercury sulfate)
• ORP conversion to SHE
• Calculate rH with factor
Delta function
When a delta value is entered, the
system calculates the difference
Output value = measured value – delta value
The output value controls all outputs
and is shown on the display. When
the delta function has been activated
simultaneously with temperature
compensation, the temperature is
compensated first and then the delta
value is subtracted.
When delta function is switched
on, “∆” appears in the display in
measuring mode.
ORP/rH value (Administrator)
24.0 °C
Abort
Ref el
7.00 pH
Ag/AgCl, KCl 1m
OK
ORP conversion
rH with factor
Ag/AgCl, KCl 3m
Hg,Ti/TiCl,KCl 3.5m
Hg/Hg2SO4,K2SO4 sat
Outp I1 5.70 mA
Favorites menu
∆
Delta function (Administrator)
24.0 °C
Abort
Delta function
7.00 pH
pH
OK
Delta value
mV
ORP
rH
Off
pH
°C
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.
Содержание Protos II 4400
Страница 18: ...18 ...
Страница 23: ...23 Wiring Examples Connection of ISM sensor Transparent Black PH 3400 X 035 ...
Страница 73: ...73 ...
Страница 88: ...88 Appendix Buffer table Mettler Toledo ...
Страница 90: ...90 Appendix Buffer table DIN 19267 ...
Страница 92: ...92 Appendix Buffer table Techn buffers to NIST ...
Страница 93: ...93 Appendix Buffer table Hamilton ...
Страница 94: ...94 Appendix Buffer table Kraft ...
Страница 95: ...95 Appendix Buffer table Hamilton A ...
Страница 96: ...96 Appendix Buffer table Hamilton B ...
Страница 97: ...97 Appendix Buffer table HACH ...
Страница 100: ...100 ...
Страница 111: ...111 Index Protos PH 3400 X 035 Module ...