36
Data entry of premeasured electrodes
Entry of values for zero point, slope and isothermal potential of a pH electrode.
The values must be known, e.g. determined beforehand in the laboratory.
NOTICE!
Input of an isothermal potential V
iso
also applies to the calibration
methods
• Calimatic
• Manual input
• Product calibration
For an explanation of the isothermal potential, refer to Pg 37.
During calibration the module is in function check (HOLD) mode. Current
outputs and relay contacts of the module behave as configured (BASE).
Menu Display
Data entry
Select: Data entry of premeasured
electrodes
Remove electrode and connect
premeasured electrode.
Open “Sensor replacement”.
Enter the values for
• Zero
• Slope
• Isothermal potential
Return using softkey or press meas to
return to measurement.
Data entry
Return
When changing sensors perform
First cal for statistics!
25.6°C
7.00 pH
+07.00pH
Zero
Slope
Isothermal pot0000 mV
058.0 mV/pH
Calibration / Adjustment
Calibration by entering data from premeasured electrodes
Sensor replacement
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.
Содержание Protos II 4400
Страница 18: ...18 ...
Страница 23: ...23 Wiring Examples Connection of ISM sensor Transparent Black PH 3400 X 035 ...
Страница 73: ...73 ...
Страница 88: ...88 Appendix Buffer table Mettler Toledo ...
Страница 90: ...90 Appendix Buffer table DIN 19267 ...
Страница 92: ...92 Appendix Buffer table Techn buffers to NIST ...
Страница 93: ...93 Appendix Buffer table Hamilton ...
Страница 94: ...94 Appendix Buffer table Kraft ...
Страница 95: ...95 Appendix Buffer table Hamilton A ...
Страница 96: ...96 Appendix Buffer table Hamilton B ...
Страница 97: ...97 Appendix Buffer table HACH ...
Страница 100: ...100 ...
Страница 111: ...111 Index Protos PH 3400 X 035 Module ...