29
Menu Display
Select calibration method (pH)
Open calibration
Press menu key to select menu.
Select calibration using arrow keys,
press enter to confirm, passcode 1147
(To change passcode, select:
Parameter setting/System control/
Passcode entry).
Calibration:
Select “Module PH”
Select calibration method:
• Automatic buffer recognition
• Manual entry of buffer values
• Product calibration
(Calibration with sampling)
• Entry of previously measured electrode data
• ORP calbration/adjustment
• Temp probe adjustment (with Protos II 4400(X))
Menu selection
Return to meas
7.00 pH
25.6 °C
Select:
[enter]
Lingua
Calibration
25.6 °C
Return
Info
7.00 pH
Module PH 3400-035
Module PH 3400-035
25.6 °C
Return
Data entry: premeasured electrodes
Manual: entry of buffer values
Product calibration
7.00 pH
Calimatic: automatic calibration
ORP calibration
When you open the Calibration
menu, the analyzer automatically
proposes the previous calibration
method. If you do not want to
calibrate, press the ”Return” softkey or
the meas key.
During calibration the module is in
function check (HOLD) mode.
Current outputs and relay contacts
of the module behave as configured
(Module BASE).
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.
Содержание Protos II 4400
Страница 18: ...18 ...
Страница 23: ...23 Wiring Examples Connection of ISM sensor Transparent Black PH 3400 X 035 ...
Страница 73: ...73 ...
Страница 88: ...88 Appendix Buffer table Mettler Toledo ...
Страница 90: ...90 Appendix Buffer table DIN 19267 ...
Страница 92: ...92 Appendix Buffer table Techn buffers to NIST ...
Страница 93: ...93 Appendix Buffer table Hamilton ...
Страница 94: ...94 Appendix Buffer table Kraft ...
Страница 95: ...95 Appendix Buffer table Hamilton A ...
Страница 96: ...96 Appendix Buffer table Hamilton B ...
Страница 97: ...97 Appendix Buffer table HACH ...
Страница 100: ...100 ...
Страница 111: ...111 Index Protos PH 3400 X 035 Module ...