Sensoface
Sensoface is a graphic indication of the sensor condition.
Sensocheck
Automatic monitoring of glass and reference electrode
Parameter
Slope
Zero
Reference impedance
Glass impedance
Response time
Fine
Standard
Coarse
Calibration timer
Sensor wear
Sensoface Criteria
Standard*
59.2
7.00
Rcal **
Rcal **
Critical range
< 53.3 or > 61
< 6.00 or > 8.00
< 0.6 Rcal or > 100 kΩ+ 0.5 Rcal
< 0.3 Rcal or > 3.5 Rcal
120 sec
80 sec
60 sec
when 80 % expired
as specified (ISM sensors only)
* Applies to standard electrodes with pH = 7.00
** Rcal is determined during calibration
The “smileys” provide information on wear
and required maintenance of the sensor
(“friendly” - “neutral” - “sad”).