Servicing
Vacuum pump system SCC 950
46
Translation of original Operating Instructions, English, KNF 304771-304773 07/18
9.3.7. Final steps
1. Place the pump back into the vacuum pump system and tight-
en the four attachment screws (Fig. 18/
9
) hand tight.
2. Replace the connections at the suction and pressure sides of
the pump:
To do this, place the two connection fittings (
8
) in the connec-
tion blocks and then fix them in place with the two spring steel
clips (
7
).
3. Replace plugs
3
,
4
,
5
and
6
on the electrical connection of the
pump.
4. Put the grip assembly (
2
) back onto the vacuum pump system
and tighten the four attachment screws (
1
) hand tight.
5. Reconnect suction and pressure lines to the vacuum pump
system.
6. If applicable, fix the charging station to the vacuum pump
system again (see Fig. 8, p. 21).
7. Reconnect the vacuum pump system to the mains power.
If you have any questions about servicing, call your KNF technical
adviser (see last page for contact telephone number).