Vacuum pump system SCC 950
Servicing
Translation of original Operating Instructions, English, KNF 304771-304773 07/18
39
9.3. Replacing the diaphragms and valve plates
Vacuum pump system disconnected from mains power
and de-energized
Vacuum pump system is clean and free of hazardous ma-
terials.
Hoses removed from pneumatic system inlet and outlet.
Charging station not attached to vacuum pump system
(remove it beforehand)
Qty Material
1
Phillips screwdriver no. 2
1
Allen wrench 3
1
Allen wrench 2.5
1
Open-end wrench, width 7 mm
1
Spare parts set (see Chapter 11, page 52)
1
Pencil
Tab. 8
Always replace diaphragms and valve plates together to main-
tain the pump capacity.
WARNING
Health hazard due to dangerous substances in the
vacuum pump system and pump
Depending on the medium transferred, caustic burns
or poisoning are possible.
Wear protective clothing if necessary, e.g.
protective gloves.
Flush the vacuum pump system before replacing
the diaphragms and valve plates (see Chapter
9.2.1, page 37).
CAUTION
Danger of burns from hot pump parts
The pump head or motor may be hot even after the
pump has been shut off.
Allow the pump to cool off after operation.
Requirements
Material and tools
Information on procedure