Vacuum pump system SCC 950
Troubleshooting
Translation of original Operating Instructions, English, KNF 304771-304773 07/18
47
10. Troubleshooting
DANGER
Risk of electric shock, danger of death
Disconnect the pump from the power supply
before working on the pump.
Make sure that the pump is de-energized.
Vacuum pump system generally: see Tab. 9.
Hand terminal: see Tab. 10.
Software operation: see Tab. 11.
Vacuum pump system generally
Error
Cause
Remedy
The vacuum pump system
is switched on, but the
power switch does not
illuminate.
Power cable not plugged
in.
Plug the vacuum pump system's
mains power cable into a properly
installed grounded socket.
No voltage in the mains.
Check the room's fuses.
Fuses in vacuum pump
system are blown.
1. Identify and eliminate cause of
overload.
2. Replace the mains fuse(s) of the
vacuum pump system:
At the fuse drawer
(Fig. 2/
17
, p. 13) press the lock-
ing strap downward and remove
the fuse drawer.
Replace defective fuse(s)
(For fuse specifications, refer to
Chapter 4.1, page 9f; for the or-
der number of the fuse, see
Chapter 11, page 52).
Slide the fuse drawer in until it
engages.
Insufficient vacuum gener-
ated although the pump is
running.
Vacuum chamber leaking
1. Check by closing the gas inlets
17
and
18,
p. 13). If the
pump achieves sufficient vacuum,
the vacuum chamber has a leak.
2. Repair the leak in the vacuum
chamber.
The O-ring at the holder for
the suction side collection
flask is not installed cor-
rectly.
Align the O-ring
If the O-ring is defective, replace it
(order no. see Chapter 11, page 52).
Hose connection leaking.
Make sure the hoses are properly
seated on the hose connectors.
Replace leaking hoses.
Replace leaking hose connectors.
For the order number, refer to Chap-
Tab. 9 (1st part)