Vacuum pump system SCC 950
Technical data
Translation of original Operating Instructions, English, KNF 304771-304773 07/18
11
4.2. Hand terminal of the vacuum pump system
Dimensions: W x H x D [mm]
91 x 190 x 65
Weight [kg]
0.5
Operating voltage [V DC]
24
Operating current [A]
1
Wireless connection's frequen-
cy band [GHz]
2.4
Wireless range
Approx. 50 m without obstacles
Approx. 10 m through walls
Power supply
Via integrated batteries or
supplied power supply unit;
cable length approx. 1.50 m,
with adapter set:
EURO / UK / USA / AUS
DC charging socket
External diameter: 6.3 mm
Inside diameter: 2 mm
Batteries
4 x AA 1.2 V 2600 mAh; quick-
charge capable; see spare
parts list in Chapter 11
Battery service life*
Up to 12 hours, depending on
number of entries and data
transmission
Charging time*
about 7 h
Tab. 3
* Applies to standard included batteries
When charging the hand terminals of the SCC 950 vacuum
pump system, use only the original power supply from KNF.
Several SCC 950 vacuum pump systems can be operated
simultaneously within the range of the wireless connection
using the associated hand terminals.
The wireless connection between the hand terminal and the
SCC 950 vacuum pump system is robustly compatible with cell
phones and Bluetooth devices in the immediate area.