Vacuum pump system SCC 950
Operation
Translation of original Operating Instructions, English, KNF 304771-304773 07/18
23
7. Operation
7.1. Initial start-up
Before switching on the vacuum pump system, check the following
points:
Prerequisites for start-up
All hoses attached properly
Fan openings not blocked
Specifications of the power supply correspond with the
data on the vacuum pump system’s type plate
Recirculating cooler or cold water connection ready on
high-performance condenser
The high-performance condenser's gas outlet is not
blocked (high-performance condenser is not pressure-
proof)
Vacuum pump system at ambient temperature
Ensure that hand terminals are properly matched up with
the appropriate vacuum pump system
If the vacuum system is ventilated through the air inlets
(Fig. 2/
17
and
18
, p. 13) no reactive, explosive, or other-
wise dangerous mixtures can form (use inert gas if nec-
essary)
When the gas ballast valve is open (blue LED lit), no re-
active, explosive, or otherwise hazardous mixtures can
form (otherwise, contact KNF Service).
Tab. 4
Operate the vacuum pump system only with the operating
parameters and conditions described in Chapter 4, Technical
Data (page 9).
Make sure the vacuum pump system is used properly (see
Exclude the possibility of the vacuum pump system being used
improperly (see Chapter 2.2, page 6).