Vacuum pump system SCC 950
Setup and connection
Translation of original Operating Instructions, English, KNF 304771-304773 07/18
19
6. Setup and connection
Connect the vacuum pump system only under the operating
parameters and conditions described in Chapter 4. Technical
data (page 9).
Observe the safety precautions (see Chapter 3, page 7).
6.1. Installation
Before installation, allow the vacuum pump system to reach
ambient temperature at the installation location.
Install the vacuum pump system so that the two fans are not
blocked.
Make sure that the installation location is dry and that the
vacuum pump system is protected against water in the form of
rain, spray, splashes and drips.
Choose a safe location (flat surface) for the vacuum pump
system.
Protect the vacuum pump system from dust.
Protect the vacuum pump system from vibration and impact.
Only connect components to the vacuum pump system that
are designed for the pneumatic data of the vacuum pump sys-
tem (see Chapter 4, page 9).
A recirculating cooler or flowing cold water (or another cooling
medium) is needed to cool the high-performance condenser to the
condensation temperature.
Hand terminal 1 of the SCC 950 vacuum pump system is secured
at the factory in order to prevent damage during transportation.
The transportation safeguard must be disengaged before the hand
terminal can be removed. To do this, rotate the knurled knob (Fig.
2/
2
, page 13) out until the hand terminal can be removed.
The transportation safeguard can be screwed back in before
transporting the vacuum pump system in the future.
Install the separately included collection flasks (Fig. 2/
8
and
9
,
page 13) using the included flask clips (Fig. 2/
7
, page 13).
Set up the charging station for hand terminal 2 (see Fig.
6, page 20).
Basically, there are four different setup or installation options:
1.) Bench version 1 (see Fig. 6, top)
2.) Bench version 2 (see Fig. 6, center)
3.) Wall installation (see Fig. 6, bottom)
4.) Attachment to the vacuum pump system (see Fig. 6, bot-
tom)
Cooling air supply
Installation location
Connected
components
Coolant for
high-performance condenser
Remove transportation safe-
guard
Installing the collection flask
Set up the charging station