Operating the vacuum pump system
Vacuum pump system SCC 950
32
Translation of original Operating Instructions, English, KNF 304771-304773 07/18
1.) Start Automatic operat-
ing mode
2.) Boiling point detected;
system regulates to this
pressure
3.) Automatic pressure re-
duction
Fig. 14: Automatic function sequence
If you press the arrow key (see Fig. 11/
7
) during automatic
pressure reduction, the current pressure is applied as the new
setpoint pressure and the vacuum pump system regulates to
this pressure. The current setpoint pressure is shown in the
display in place of the arrow.
If you press again on the setpoint pressure value, automatic
pressure reduction is reactivated and the arrow appears in the
display after about 25 seconds.
Entering values in Function operating mode
The desired pressure curve is entered via data points that connect
the vacuum pump system to pressure ramps.
Fig. 15: Menu in Function operating mode
Entering the data points (up to 12):
1. Length of time
from previous data point: ∆ t
2. Setpoint pressure: p (mbar) / p (bar) / p (hPa) / p (Torr)