P 2.1L user manual V6840, version 3.6
30
Troubleshooting
Line in time table is
empty
Edit the program line.
Link is loaded
First unload the link then change the link or
delete it.
Link is running
Wait until the link has been completed, then
change the link or delete it.
M
Max. flow limit
reached
Confirm, pump continues running
Maximum pres-
sure! System
stopped
Reduce the pressure or adjust the upper pres-
sure limit.
Restart the system
Minimum pressure!
System stopped
Increase the pressure or adjust the lower
pressure limit.
Restart the system
Motor failure: max
current
Switch the device off and on. Inform the techni-
cal support of the manufacturer in case the sys-
tem message repeats itself.
N
No gradient is avail-
able in isocratic
mode
Change the
setup
settings or change the gradi-
ent in the program or in
setup
.
No link available
Create a link and edit it.
No link available.
Pls edit link first
Create a link and edit it.
No time table to
start
Edit the data by means of the chromatography
software.
Non-existing com-
ponent is set to
non-0 value
Switch on the channel or edit the data using the
chromatography software.
P
Program does not
exist
Create a program.
Program is running
Quit program or wait until program has been
completed.
Program not com-
patible with pump
head
Modify the program or replace the pump head.
S
Sum of compo-
nents is not 100
Change the entry.
T
This link is used in
WAKEUP
First quit or delete wakeup program (wu = Wake
Up), then edit or delete link.
This program is
used in a link
First pause or delete the link, then edit or delete
data by means of the chromatography software.
This program is
used in WAKEUP
First quit or delete wakeup program (wu = Wake
Up), then edit or delete data by means of the
chromatography software.
System message
Solution