P 2.1L user manual V6840, version 3.6
16
Installation
6OC
TTL output
Levels:
passive 0 V
active 5 V
Pulse:
5 V for at least 1000 ms
GND
Reference point of the voltage at the signal inputs.
Relay contact
The contact is on a floating basis. Its setting depends
on the settings in the Control Unit or software.
Steady-rate signal:
passive = open relay contact
active = closed relay contact
Pulse:
Closed relay contact for at least 1000 ms
Permissible load of the relay contact: 1 A/ 24 V
DC
Relay contact
The contact is on a floating basis. Its setting depends
on the settings in the Control Unit or software.
Steady-rate signal:
passive = open relay contact
active = closed relay contact
Pulse:
Closed relay contact for at least 1000 ms
Permissible load of the relay contact: 1 A/ 24 V
DC
Analog GND
Reference point of the voltage at the signal inputs.
Analog out
Voltage range 0 – 5 V, scalable
Analog in
Voltage range 0 – 10 V
10 V according to maximum flow rate
Start GND
Reference point of the voltage at the signal inputs.
Start OUT
TTL output
Levels:
passive 5 V
active 0 V
Start IN
TTL input
Low active
Secure switching threshold at least 10 mA
After receiving a signal (short-circuit to ground) from
an external device, the device starts. If controlled with
software, an electronic trigger is send through the
LAN.
Assignment
Connection
Function