P 2.1L user manual V6840, version 3.6
Installation
17
Connecting Cables to the Terminal Strip
To control one device through another, you use the multi-pin connector.
To use remote control, you have to connect cables to the terminal strip
(both included with delivery). The single ports are used to exchange con-
trol signals.
Prerequisite
The device has been turned off.
The power plug has been pulled.
Tools
Operating tool
Electronic defect
Connecting cables to the multi-pin connector of a switched on device
causes a short circuit.
Turn off the device before connecting cables.
Pull the power plug.
Electronic defect
Electrostatic discharge can destroy the electronics.
Wear a protective bracelet against electrostatic discharge and ground.
Process
1. Push the operating tool
3
into an
upper small opening on the front
of the terminal strip
1
.
2. Lead the cable into the opening
2
below the inserted operating
tool.
3. Remove the operating tool.
Error OUT
TTL output
Levels:
passive 5 V
active 0 V
Error IN
TTL input
Low active
Secure switching threshold at least 10 mA
After receiving a signal (short-circuit to ground) from
an external device, an error message appears and the
device stops.
Error GND
Reference point of the voltage at the signal inputs.
+24V
Event-controlled switching of 24 V
against GND
Protection: 24 V – 200 mA
+5V
Provides a voltage of 5 V with respect to GND. This
makes it possible to supply a consumer that is
switched by an EVENT.
Protection: 5 V – 50 mA
GND
Reference point of the voltage at the signal inputs.
Assignment
Connection
Function
2
3
1