P 2.1L user manual V6840, version 3.6
12
Installation
Next steps
Connect the plug of the valve block.
Bleeding the Pump
Prerequisite
The capillaries have been connected.
Tool
Syringe
Column defect
Damage to the column due to bleeding.
Open the venting screw.
Remove the column.
Note:
Before the pump can be used, it must be bleeded.
Result
No air bubbles in the pump head and in the capillaries.
Leak Management
The leak management consists of the leak sensor and the drainage system
(funnels, hoses, nozzles). The drainage system ensures that escaping liq-
uids flow into a waste bottle. When leaks are registered by the leak sen-
sor, the LED flashes red. Both the device and the data acquisition via
chromatography software are stopped.
Prerequisite
The front cover has been removed.
3. Tighten the fastening screw by
hand.
4. Plug a cap fitting into the
unused inlet.
Fig. 18
Inlets of the valve block
4
Procedure
Process
Figure
1. Open the venting screw
1
of
the pressure sensor.
2. With the syringe, extract fluid
through the bleed port
2
.
3. If the extracted fluid flows con-
tinuously, stop suction and
close the venting screw.
Fig. 19
Venting screw of the pres-
sure sensor
1
2
Process
Figure
1. Carefully push the funnel
1
into the center opening of the
capillary guide
2
.
Fig. 20
Funnel and capillary guide
1 2