![Knauer Azura P 2.1L Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/knauer/azura-p-2-1l/azura-p-2-1l_user-manual_1979875008.webp)
P 2.1L user manual V6840, version 3.6
Safety for Users
5
Accident prevention regulations published by the accident insurance
companies for laboratory work
Filtration of substances under analysis
Use of inline filters
Once the capillaries have been used, never re-use them in other areas
of the HPLC system.
Only use a given PEEK fitting for one specific port and never re-use it
for other ports. Always install new PEEK fittings on each separate port.
Follow KNAUER or manufacturer’s instructions on caring for the
colums.
More safety-relevant information is listed below:
flammability: Organic solvents are highly flammable. Since capillaries
can detach from their screw fittings and allow solvent to escape, it is
prohibited to have any open flames near the analytical system.
solvent tray: Risk of electrical shock or short circuit if liquids get into the
device's interior. For this reason, place all bottles in a solvent tray.
solvent lines: Install capillaries and tubing in such a way that liquids
cannot get into the interior in case of a leak.
leaks: Regularly check if any system components are leaking.
power cable: Defective power cables are not to be used to connect the
device and the power supply system.
self-ignition point: Only use eluents that have a self-ignition point
higher than 150 °C under normal ambient conditions.
power strip: If several devices are connected to one power strip, always
consider the maximum power consumption of each device.
power supply: Only connect devices to voltage sources, whose voltage
equals the device's voltage.
toxicity: Organic eluents are toxic above a certain concentration.
Ensure that work areas are always well-ventilated! Wear protective
gloves and safety glasses when working on the device!
Where is use of the device prohibited?
Never use the system in potentially explosive atmospheres without appro-
priate protective equipment. For further information, contact the Techni-
cal Support of KNAUER.
Secure decommissioning
Take the device completely out of operation by either switching off the
power switch or by pulling the power plug.
Opening the device
The device may be opened by the KNAUER Technical Support or any
company authorized by KNAUER only.
Warning notifications
Possible dangers related to the device are divided into personal and
material damage in these instructions.
DANGER indicates a hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING indicates a hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION indicates a hazardous situation which, if not
avoided, could result in minor or moderate injury.
NOTICE is used to address practices not related to phys-
ical injury.
Содержание Azura P 2.1L
Страница 1: ...Pump P 2 1L User Manual V6840 HPLC...